首页> 外文会议>SPIE Conference on Frontiers in Biological Detection: From Nanosensors to Systems >Optical Waveguide Biosensor Based on Cascaded Mach-Zehnder Interferometer and Ring Resonator with Vernier Effect
【24h】

Optical Waveguide Biosensor Based on Cascaded Mach-Zehnder Interferometer and Ring Resonator with Vernier Effect

机译:基于级联Mach-Zehnder干涉仪的光波导生物传感器和vernier效应的环谐振器

获取原文

摘要

Optical waveguide biosensors based on silicon-on-insulator (SOI) have been extensively investigated owing to its various advantages and many potential applications. In this article, we demonstrate a novel highly sensitive biosensor based on cascaded Mach-Zehnder interferometer (MZI) and ring resonator with the Vernier effect using wavelength interrogation. The experimental results show that the sensitivity reached 1,960 nm/RIU and 19,100 nm/RIU for sensors based on MZI alone and cascaded MZI-ring with Vernier effect, respectively. A biosensing application was also demonstrated by monitoring the interaction between goat and antigoat immunoglobulin G (IgG) pairs. This integrated high sensitivity biosensor has great potential for medical diagnostic applications.
机译:由于其各种优点和许多潜在应用,基于绝缘体(SOI)的光波导生物传感器已被广泛研究。在本文中,我们展示了一种基于级联Mach-Zehnder干涉仪(MZI)的高度敏感的生物传感器和使用波长询问的vernier效应的环谐振器。实验结果表明,基于单独的MZI和级联的MZI环,敏感性达到1,960nm / Riu和19,100nm / Riu,分别与vernier效应级联。还通过监测山羊和抗抗煤免疫球蛋白G(IgG)对的相互作用来证明生物沉积应用。这种集成的高灵敏度生物传感器具有很大的医疗诊断应用潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号