首页> 外文会议>Conference on optical coherence tomography and coherence techniques >A method to calibrate phase fluctuation in polarization-sensitive swept-source optical coherence tomography
【24h】

A method to calibrate phase fluctuation in polarization-sensitive swept-source optical coherence tomography

机译:一种校准极化敏感扫描源光学相干断层扫描中相波动的方法

获取原文

摘要

A phase fluctuation calibration method is presented for polarization-sensitive swept-source optical coherence tomography (PS-SS-OCT) using continuous polarization modulation. The method consists of the generation of a continuous triggered tone-burst waveform rather than an asynchronous waveform by use of a function generator and the removal of the global phases of the measured Jones matrices by use of matrix normalization. This could remove the use of auxiliary optical components for the phase fluctuation compensation in the system, which reduces the system complexity. Phase fluctuation calibration is necessary to obtain the reference Jones matrix by averaging the measured Jones matrices at sample surfaces. Measurements on an equine tendon sample were made by the PS-SS-OCT system to validate the proposed method
机译:介绍了使用连续偏振调制的偏振敏感扫描源光学相干断层扫描(PS-SS-OCT)的相位波动校准方法。该方法包括通过使用函数发生器来产生连续触发的色调突发波形而不是异步波形,并通过使用矩阵标准化来移除测量的琼斯矩阵的全局阶段。这可以消除使用辅助光学元件在系统中的相位波动补偿,这降低了系统复杂性。通过在样品表面上平均测量的琼斯矩阵来获得相位波动校准以获得参考琼斯矩阵。通过PS-SS-OCT系统进行马腱样品的测量以验证所提出的方法