Germanium Detector Calibration According to the Standard NF M 60-810 without Using Radioactive Sources

机译:锗探测器根据标准NF M 60-810校准而不使用放射源

获取原文

摘要

In-situ gamma ray spectrometry is used to determine the specific activities of natural and artificial radioactive nuclides in the soil with a good accuracy. This method is very interesting for environmental measurements and lead of soil determination activity. It is a cheaper method than analysis of great amounts of soil samples in the laboratory. As there is not standard soil, detection efficiency can be estimated using either statistical tools or combination of radioactive point sources calibration thanks to mathematical models of NF M 60-810 standard representing the radionuclide distribution in soil. Experimental determination of detection efficiency requires a large number of operations involving the handling of radioactive standards in the energy range from 0.06-2 MeV. For these reasons, detection efficiency model has been determinate without using radioactive sources.
机译:原位伽马光线光谱法用于确定土壤中的天然和人造放射性核素的特定活动,具有良好的准确性。这种方法对于环境测量和土壤测定活动的铅非常有趣。它是一种更便宜的方法,而不是实验室中的大量土壤样品的分析。由于没有标准的土壤,由于NF M 60-810标准的数学模型,可以使用统计工具或放射点源校准的组合来估计检测效率,这是NF M 60-810标准的数学模型代表土壤中的放射性核素分布。检测效率的实验确定需要大量的操作,涉及在0.06-2 meV的能量范围内处理放射性标准的处理。由于这些原因,检测效率模型已经确定而不使用放射源。