首页> 外文会议>Frontiers in Biomedical Devices Conference >FINITE ELEMENT MODELING OF THE LEFT ATRIUM TO FACILITATE THE DESIGN OF AN ENDOSCOPIC ATRIAL RETRACTOR
【24h】

FINITE ELEMENT MODELING OF THE LEFT ATRIUM TO FACILITATE THE DESIGN OF AN ENDOSCOPIC ATRIAL RETRACTOR

机译:左心房的有限元建模,便于内窥镜心房牵开器设计

获取原文

摘要

A finite element model of the left atrium, incorporating detailed anatomical features and realistic material characteristics, was built to investigate the interaction of heart tissue and surgical instruments. This model was used to facilitate the design of an endoscopically deployable atrial retractor for use in minimally invasive, robotically assisted (MIRA) mitral valve repair. The left atrial geometry was imported directly from MRI data of an explanted porcine heart, and material properties were derived from experimental testing of cardiac tissues. Model accuracy was verified by comparing simulated cardiac wall deflections to those measured by MRI. Finite element analysis was shown to be an effective tool for analyzing instrument/tissue interactions and for designing surgical instruments.
机译:建立了一种有限元模型,包括详细解剖特征和现实材料特性,以研究心脏组织和手术器械的相互作用。该模型用于促进内窥镜可展开的心房牵开器的设计,用于微创,机器人辅助(Mira)二尖瓣修复。左心房几何形状直接从外植入的猪心脏的MRI数据进行,而材料特性来自于心脏组织的实验试验。通过将模拟的心脏壁偏转与MRI测量的那些进行比较来验证模型精度。有限元分析显示为分析仪器/组织相互作用和设计手术器械的有效工具。