Application of time sampling in brain CT perfusion imaging for dose reduction

机译:时间抽样在脑CT灌注成像中的应用剂量减少

获取原文

摘要

The purpose of this study is to determine a stable sampling rate not to be affected by sampling shift for reducing radiation exposure with time sampling and interpolation in cerebral perfusion CT examination. Original images were obtained every 1 second for 40 time series from 3 patients, respectively. Time sampling was performed with sampling intervals (SI) from 2 to 10 seconds. Sampling shift was applied from +1 to SI-1 for each sampling rate. For each patient, 30 tissue concentration time-course data were collected, and arterial input curves were fitted by gamma-variate function. The sinc function was introduced for interpolation. Deconvolution analysis based on SVD was performed for quantifying perfusion parameters. The perfusion values through time-varying sampling and interpolation were statistically compared with the original perfusion values. The mean CBF values with increase of sampling interval and shift magnitude from the collected data had a wider fluctuation pattern centering around the original mean CBF. The mean CBV values had a similar tendency to the mean CBF values, but a relatively narrower deviation. The mean MTT values were fluctuated reversely to the trend of the mean CBF values. The stable sampling interval for quantifying perfusion parameters with lower radiation exposure was statistically acceptable up to 4 seconds. These results indicate that sampling shift limits sampling rate for acquiring acceptable perfusion values. This study will help in selecting more reasonable sampling rate for low-radiation-dose CT examination.
机译:本研究的目的是确定不受采样转移的稳定采样率,以减少辐射暴露在脑灌注CT检查中的时间采样和插值。每1秒获得原始图像,分别从3名患者40次序列获得。以2至10秒的采样间隔(Si)进行时间采样。为每个采样率从+1到Si-1施加采样换档。对于每位患者,收集了30个组织浓度时间过程数据,并且通过伽马变化功能装配动脉输入曲线。引入了Interpolation的SIST功能。基于SVD的去卷积分析进行定量灌注参数。通过时变采样和插值通过时变采样和插值与原始灌注值进行统治性值。根据所收集的数据的采样间隔和移位幅度的增加,平均CBF值在原始平均CBF周围具有较宽的波动图案。平均CBV值具有与平均CBF值相似的趋势,而是相对较窄的偏差。平均MTT值反向波动到平均CBF值的趋势。用于量化具有较低辐射暴露的灌注参数的稳定采样间隔在统计学上可接受至4秒。这些结果表明采样移位限制采样率以获取可接受的灌注值。本研究将有助于为低辐射剂量CT检查选择更合理的采样率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号