The wavelength-dependent characteristic of the floating-reference point in the non-invasive glucose sensing

机译:非侵入性葡萄糖感应中的浮选点的波长依赖性特征

获取原文

摘要

In the noninvasive blood glucose sensing using near-infrared spectroscopy, the physiological noise was one of the biggest challenges. In order to efficiently reduce the influence of the physiological background variations on the diffuse reflectance spectra, the floating-reference method was used by differentially processing two signals from reference point and measuring point. In this paper, the wavelength-dependent characteristic of the floating-reference point was discussed by simulation and primary experiment. First, the wavelength-dependent characteristic of intralipid- 5% solution was investigated in the wavelength range of 1300-1600nm. And source-detector distance for reference point in the wavelength of 1300nm was conducted by different concentration of scatter media including 2%, 5% and 10% intralipid solution. Then the single-layer and three layers skin model were built to investigate the wavelength characteristic of reference point. The water displacement coefficients and relative large change in glucose concentration were considered in the simulation. Finally, the primary experiment of intralipid model was conducted to validate the wavelength dependence of reference point. The result showed that, the floating reference will not exist in the strong absorption region (near 1450nm) and the region where the change of absorption coefficient is positive (high than 1525nm) due to the corporate influence of scattering and absorption coefficient. And the wavelength-dependent characteristic is consistent for intralipid solution and the skin model.
机译:在使用近红外光谱的非血糖血糖感测中,生理噪音是最大的挑战之一。为了有效地降低漫反射光谱对漫反射谱的生理背景变化的影响,通过差异地处理来自参考点和测量点的两个信号来使用浮选参考方法。本文通过模拟和初级实验讨论了浮选点的波长依赖性特性。首先,在1300-1600nm的波长范围内研究了含有伏特拉里哌啶-5%溶液的波长依赖性特性。在1300nm波长的参考点的源检测器距离由不同浓度的散射培养基进行,包括2%,5%和10%intraliPid溶液。然后建立单层和三层皮肤模型以研究参考点的波长特性。在模拟中考虑了葡萄糖浓度的水分位移系数和相对大的变化。最后,进行了intralipid模型的主要实验,以验证参考点的波长依赖性。结果表明,由于散射和吸收系数的企业影响,强吸收区域(接近1450nm)和吸收系数变化为正(高于1525nm)的区域,不存在浮动参考。并且波长依赖性特性对于肝癌溶液和皮肤模型一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号