Experimental Verification of Simplified Loading Pattern Employed in Calibration of Force Measuring Devices

机译:力测量装置校准中采用简化装载模式的实验验证

获取原文

摘要

Calibration of force measuring devices according to ISO 376 requires several hours, owing to the requirement to perform numerous preloading cycles. In view of these circumstances, an abbreviated loading pattern was proposed for shortening the time required for a calibration operation. Experiments were performed in order to compare the results of calibration by the ordinary loading pattern and those by the abbreviated calibration. Slight differences in output of the force measuring devices were observed; the differences were small enough in relation to the uncertainty of the calibrated values that they did not affect classifications of the force measuring devices.
机译:由于要求执行许多预加载循环,根据ISO 376的力测量装置的校准需要几个小时。鉴于这些情况,提出了一种缩写的加载模式,用于缩短校准操作所需的时间。进行实验,以比较普通装载模式的校准结果和通过缩写校准的结果。观察到力测量装置的输出的轻微差异;与它们不影响力测量装置的分类的校准值的不确定性,差异足够小。