Spectral performance of mercuric iodide gamma-ray detectors at elevated temperatures

机译:升高温度下汞碘化物γ射线探测器的光谱性能

获取原文

摘要

The effects of elevated temperatures on the spectral performance of a planar mercuric iodide (HgI_2) gamma ray detector were evaluated at 25°C, 35°C, 45°C and 55°C using two test isotopes. ~(137)Cs at 662 keV and ~(241)Am at 59 keV. Spectral analysis was used to determine the spectral parameters (i.e. %FWHM of the full energy peak, the peak channel position and the peak to background ratio). Spectral performance degraded slightly with increasing test temperatures, but recovered on returning to ambient conditions. The results demonstrate that temperature excursions up to 55°CC minimally degrade the spectral performance of mercuric iodide detectors.
机译:使用两种测试同位素在25℃,35℃,45℃和55℃下评价升高的温度对平面酰胺碘化物(HGI_2)γ射线检测器的光谱性能的影响。 〜(137)在662 kev和〜(241)上午59 kev。光谱分析用于确定光谱参数(即全能峰值的%FWHM,峰值通道位置和峰值与背景比)。光谱性能略微降解,随着测试温度的增加,但在恢复到环境条件下恢复。结果表明,高达55°CC的温度偏移最小地降低了羊毛碘化物探测器的光谱性能。