首页> 外文会议>Optical coherence tomography and coherence techniques conference >Precision of quantitative absorption measurement with spectroscopic optical coherence tomography
【24h】

Precision of quantitative absorption measurement with spectroscopic optical coherence tomography

机译:用光谱光学相干断层扫描的定量吸收测量精度

获取原文

摘要

A state of the art TiAl_2O_3 laser (λ = 780 nm, Δλ = 170nm, P_(out) = 350 mW), interfaced to a free space OCT system was used to quantitatively extract spatially resolved absorption profiles from non-scattering phantoms. The sources of error related to instrument performance and measurement procedure were investigated. Absorption as small as 0.2mm~(-1) was measured in phantoms as thin as 180 μm with precision better than 5% over 70 nm wavelength range.
机译:用于接通到自由空间OCT系统的技术TiAl_2O_3激光(λ= 780nm,Δλ= 350mw),用于从非散射模拟定量提取空间分辨的吸收谱的定量提取空间分辨的吸收曲线。研究了与仪器性能和测量程序相关的误差来源。以小于0.2mm〜(-1)的吸收以薄为180μm,精度优于超过70nm波长范围。