Since the introduction of digital flat panel detectors into clinical routine the discussion on monitor specifications for primary soft copy reading has gained new impetus. Major concerns exist for viewing of tiny opacities such as pulmonary nodules. In this study CDRAD phantom images were acquired on a caesium iodid/amorphous silicon detector at varying exposure levels. Images were read three times by three observers on a clinical 1K and 2K monitor workstation. All typical workstation functions such as magnification and window/level setting were applied during image reading. Correct detection ratios were calculated according to the CDRAD evaluation manual. Observer ratings were highest for high dose exposure and 2K monitor reading. No significant difference was detected in the correct detection ratio of observers, however the difference between the two types of workstations (1K versus 2K monitors) despite less than 3% was significant at a 95% confidence level. This is in good accordance with recently published clinical studies, however further clinical work will be needed to strengthen this laboratory based impression. Given these subtle differences in low contrast detail detection on 1K and 2K clinical PACS workstation we should probably rethink the recommendations of various national boards for the use of 2K monitors.
机译:由于将数字平板探测器引入临床日常常规,因此对主软拷贝阅读的监测规范的讨论已经获得了新的推动力。存在用于观察肺结核等微小不透明度的主要问题。在该研究中,在不同的曝光水平下在碘碘/非晶硅检测器上获得CDRAD幻像图像。在临床1K和2K监视器工作站上的三个观察者将图像读取三次。在图像读取期间应用了所有典型工作站函数,如放大和窗口/级别设置。根据CDRAD评估手册计算正确的检测比。高剂量暴露和2K监测仪评级观察者评级最高。在观察者的正确检测比中没有检测到显着差异,但两种类型的工作站(1K与2K显示器)之间的差异尽管不到3%,但置信水平的95%显着显着。这与最近发表的临床研究有关,但是需要进一步的临床工作来加强这种基于实验室的印象。鉴于1K和2K临床PACS工作站的低对比度细节检测的这些微妙差异,我们应该重新思考各国董事会使用2K监视器的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号