首页> 外文会议>IEEE Nuclear Science Symposium Conference >An evaluation of cardiac uniformity, contrast and SNR with dual-head 180 degree and triple-head 360 degree SPECT scans
【24h】

An evaluation of cardiac uniformity, contrast and SNR with dual-head 180 degree and triple-head 360 degree SPECT scans

机译:双头180度和三头360度SPECT扫描的心脏均匀性,对比度和SNR的评估

获取原文

摘要

An experimental phantom study was performed to evaluate cardiac uniformity, contrast and SNR for two clinical cardiac SPECT imaging protocols: adjacent dual-head 180° and triple-head 360° scans. Methods: One head of a SPECT camera was used to acquire 180° and 360° projections with different times per step to simulate the clinical case where dual-head 180° and triple-head 360° each takes a total of 20 minutes. Scans were acquired with no lesion, anterior lesion and posterior lesion in the myocardium. Maximum a posteriori (MAP) reconstruction was done by Iterative Coordinate Descent (ICD) algorithm using a quadratic convex prior. The L-curve method was used to obtain the prior strength. Images both with attenuation and scatter correction (ASC) and without ASC were compared. Results: The 180° scan shows an intensity decrease in anterior apical and posterior basal regions. The 360° scan shows an intensity decrease in the posterior wall. For the anterior lesion the 180° scan has slightly better contrast while for the posterior lesion the 360° scan has slightly better contrast. Conclusions: The difference between the 180° and 360° scans is subtle and the comparison results depend on the lesion position and the view angle of the heart. A receiver operating characteristic (ROC) study of 180° vs 360° acquisition designed based on these characterizations of contrast, uniformity and noise will be necessary to evaluate overall performance.
机译:进行实验幻像研究以评估两种临床心脏SPECT成像协议的心脏均匀性,对比度和SNR:相邻双头180°和三头360°扫描。方法:SPECT摄像机的一个头用于采用每一步的不同时间进行180°和360°突起,以模拟双头180°和三头360°各自需要20分钟的临床情况。在心肌中没有病变,前病变和后病变的扫描。最大后验(MAP)重建是通过使用二次凸起的迭代坐标血统(ICD)算法完成的。 L-Curve方法用于获得先前的强度。比较了衰减和散射校正(ASC)和没有ASC的图像。结果:180°扫描显示前部和后基底区域的强度降低。 360°扫描显示后壁的强度降低。对于前部病变,180°扫描略微更好地对比,而对于后病变,360°扫描略微更好地形成鲜明对比。结论:180°和360°之间的差异是微妙的,比较结果取决于心脏的病变位置和视角。基于对比度,均匀性和噪声的比较,均匀性和噪声的设计,设计了180°Vs 360°采集的接收器操作特性(ROC)研究是为了评估整体性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号