Determining the accuracy of an eye-tracking system for laser refractive surgery

机译:确定激光屈光手术眼跟踪系统的准确性

获取原文

摘要

The accuracy of an eye tracking system designed for laser refractive surgery is determined. An enucleated porcine eye is attached to a scanning device and the movement measured using the eye tracking system. The recorded position is compared to the preprogrammed position of the scanning device. The system demonstrated an accuracy of 0.06mm for an intact cornea and 0.1mm for a cornea with a thin flap removed. The average decentration in laser refractive surgery is 0.4mm, so the implementation of this eye tracker would lead to improved alignment between the laser and eye during laser refractive surgery.
机译:确定专为激光屈光手术设计的眼睛跟踪系统的准确性。通过眼睛跟踪系统连接到扫描装置和测量的运动。将记录的位置与扫描装置的预编程位置进行比较。该系统表明,对于具有薄瓣的角膜的完整角膜和0.1mm,薄片的精度显示为0.06mm。激光屈光手术中的平均碎片为0.4mm,因此在激光屈光手术期间,该眼跟踪器的实施将导致激光和眼睛之间的改善对准。