We have constructed a link between the Physikalisch-Technische Bundesanstalt (PTB) in Braunschweig, Germany and the National Institute of Standards and Technology (NIST) in Boulder, Colorado using carrierphase Global Positioning System (GPS) receivers. The link provides a direct measurement of the frequency difference between hydrogen maser H2(PTB) and UTC(NIST), and can also be used to provide a direct comparison between the primary frequency standards at the two laboratories. Based on our previous work with this method, we expect to be able to realize the frequency comparisons with an uncertainty of about 2 x l0~(-15) using several days of averaging. This uncertainty is smaller than the combined uncertainties of the primary frequency standards in both laboratories, and it therefore supports comparison of these primary frequency standards without degrading their capabilities with the noise of the transfer system.
机译:我们在德国Braunschweig的Physikalisch-Technische Bundesanstalt(PTB)之间构建了一个联系,Colorado在科罗拉多州的博尔德国家标准和技术研究所(NIST),使用了Carriensphase全球定位系统(GPS)接收器。该链接提供了氢气发雾器H2(PTB)和UTC(NIST)之间的频率差的直接测量,并且还可用于提供两个实验室的初级频率标准之间的直接比较。基于我们之前的方法,我们希望能够使用几天的平均值来实现具有约2×10〜(-15)的不确定性的频率比较。这种不确定性小于两个实验室中初级频率标准的组合不确定性,因此支持这些初级频率标准的比较,而不会在传输系统的噪声降低其能力。