Optical tomographic scanning system for detection and monitoring

机译:用于检测和监控的光学断层扫描系统

获取原文

摘要

This optical tomography scanner for imaging (3D) applied density distribution as been evaluated for uniformity, resolution, and linearity. Signal to noise ratio of about 1000:1 in both line projection and in back projection reconstructed image has been obtained. Radiation-sensitive BANG$+R$/ polymer gels were studied using this system. Dose as low as 10 cGy up to 10 Gy for 6 MV X-ray and 6 MeV electrons from linear accelerator have been measured. Since the gels can be made UVlight sensitive, dose distribution could be used to measure exposure using this technique.
机译:该用于成像(3D)的光学断层扫描仪,用于施加密度分布,用于均匀性,分辨率和线性。已经获得了在两个线路投影和反射投影重建图像中约1000:1的信噪比。使用该系统研究了辐射敏感膨胀+ r $ /聚合物凝胶。为6 MV X射线和6MeV电子的剂量低至10cGy,从而测量来自线性加速器的6meV电子。由于凝胶可以使凝胶敏感,因此可以使用剂量分布来使用该技术测量曝光。