首页> 外文会议>Workshop on micro total analysis systems >Hydrophobic microcapillary vent for pneumatic manipulation of liquid in muTAS
【24h】

Hydrophobic microcapillary vent for pneumatic manipulation of liquid in muTAS

机译:疏水微毛细管通风口,用于液体液体的气动操纵

获取原文

摘要

For integrated multi-step muTAS, pneumatic manipulation of liquid reagents is a promising way to minimize dead volume and contamination. In such a system, passive vent valves are required for positioning and mixing of reagent droplets. This paper presents design, fabrication, and testing of a novel microstructure named Hydrophobic Microcapillary Vent (HMCV). By utilizing the negative capillary action of aqueous liquid against hydrophobic microcapillaries, one can position the liquid accurately without sensors: a water droplet of 400 pL volume was successfully positioned using air pressure up to 30 kPa. Since the HMCV has no moving parts and can be fabricated by molding technique up to 30 kPa. Since the HMCV has no moving parts and can be fabricated by molding technique at low cost, it can be built in a disposable unit of microdevices.
机译:对于集成的多步MUTAS,液体试剂的气动操纵是最小化死量和污染的有希望的方法。在这种系统中,需要无源通气阀来定位和混合试剂液滴。本文提出了一种名为疏水性微毛细管通风口(HMCV)的新型微观结构的设计,制造和测试。通过利用含水液体对疏水性微量微毛坯的负毛细管作用,可以精确地定位液体而没有传感器:使用高达30kPa的空气压力成功定位400pl体积的水滴。由于HMCV没有移动部件,并且可以通过高达30kPa的模塑技术制造。由于HMCV没有移动部件,并且可以通过成型技术以低成本制造,因此它可以内置在一次性微量的微型单元中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号