首页> 外文会议>Conference on Medical Imaging Physics of Medical Imaging >Simulation of the imaging performance of x-ray image intensifier/TV camera chains
【24h】

Simulation of the imaging performance of x-ray image intensifier/TV camera chains

机译:X射线图像增强型/电视摄像机链的成像性能模拟

获取原文

摘要

The signal and noise propagation through a generalized x-ray image intensifier/TV camera chain is modeled in terms of the modulation transfer function, two dimensional noise power spectrum, and detective quantum efficiency. The model covers effects of energy dependent x- ray absorption in the cesium iodide entrance screens, K-fluorescence and K-reabsorption, electron optics, output screens, lenses and also sensing with electronic pick-up tubes (plumbicons) or CCD sensors. Several Philips x-ray image intensifiers in combination with plumbicon TV-chains and with laboratory type CCD cameras are compared with the model. Based on the model results we further present an image simulation tool. The image simulation allows a direct evaluation of the impact of the individual components of the imaging chain on image quality. It is shown that future high resolution CCD systems can exhibit a superior image quality as compared with electronic pick-up tube systems.
机译:通过广义X射线图像增强件/电视摄像机链的信号和噪声传播在调制传递函数,二维噪声功率谱和侦探量子效率方面进行建模。该模型涵盖了碘化铯入口,K-荧光和k重吸收,电子光学,输出屏幕,透镜的能量依赖性X射线吸收的影响,以及用电子拾取管(Plumbicons)或CCD传感器感测。与模型相比,几个飞利浦X射线图像增强器与Plumbicon TV-Chains和Laboratory CCD相机相结合。基于模型结果,我们进一步呈现了一个图像仿真工具。图像仿真允许直接评估成像链的各个组分对图像质量的影响。结果表明,与电子拾取管系统相比,未来的高分辨率CCD系统可以表现出卓越的图像质量。