首页> 外文会议>IEEE Nuclear Science Symposium and Medical Imaging Conference Record >Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment
【24h】

Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment

机译:X射线CT在发射传动CT系统中使用X射线CT的衰减校正:心肌灌注评估

获取原文

摘要

We present experimental results for attenuation correction of emission tomography through use of an attenuation map derived from X-ray CT data. Both the X-ray and gamma-ray data were obtained on a prototype emission-transmission imaging system ("X-SPECT") which incorporates a third-generation X-ray CT scanner and a single-headed SPECT scanner. The two commercial imaging devices are juxtaposed such that the CT table can move the patient directly into the SPECT scanner after the X-ray scan. This enables systematic registration of SPECT and X-ray CT images. To calibrate attenuation map creation, a set of CT scans for a specific phantom geometry and X-ray technique were acquired, with varying attenuating materials in each scan. A piecewise linear fit was applied to the CT numbers and known attenuation coefficients at 140 keV, and the result is used to create attenuation maps at 140 keV from given CT images. This technique has been applied to small pigs in myocardial perfusion studies. Correcting for attenuation increased the image-derived myocardial radionuclide concentration values from 16% of the values observed for excised myocardium to 35% (ignoring background). Adding an additional correction for partial volume effects increases the image-derived value to 93% of the expected value (ignoring background). Precision, however, remains limited.
机译:我们通过使用来自X射线CT数据的衰减图,给出了衰减校正释放断层扫描的实验结果。在原型发射传输成像系统(“X-SPECT”)上获得X射线和伽马射线数据都是由第三代X射​​线CT扫描仪和单头SPECT扫描器的原型发射传输成像系统(“X-SPECT”)获得。两个商业成像装置并置,使得CT表可以在X射线扫描之后将患者直接移动到SPECT扫描仪中。这使得SPECT和X射线CT图像的系统注销。为了校准衰减地图创建,获得了一组用于特定幻像几何和X射线技术的CT扫描,每个扫描中的衰减材料变化。将分段线性配合施加到CT号和140keV的已知衰减系数,并且结果用于从给定CT图像中的140keV创建衰减图。这种技术已应用于心肌灌注研究中的小猪。校正衰减增加了从被切除的心肌观察到的值的16%的图像衍生的心肌放射核素浓度值增加到35%(忽略背景)。为部分体积效果添加额外校正将图像导出值增加到预期值的93%(忽略后台)。然而,精度仍然有限​​。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号