首页> 外文会议> >Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment
【24h】

Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment

机译:发射-透射CT系统上使用X射线CT进行SPECT的衰减校正:心肌灌注评估

获取原文

摘要

We present experimental results for attenuation correction of emission tomography through use of an attenuation map derived from X-ray CT data. Both the X-ray and gamma-ray data were obtained on a prototype emission-transmission imaging system ("X-SPECT") which incorporates a third-generation X-ray CT scanner and a single-headed SPECT scanner. The two commercial imaging devices are juxtaposed such that the CT table can move the patient directly into the SPECT scanner after the X-ray scan. This enables systematic registration of SPECT and X-ray CT images. To calibrate attenuation map creation, a set of CT scans for a specific phantom geometry and X-ray technique were acquired, with varying attenuating materials in each scan. A piecewise linear fit was applied to the CT numbers and known attenuation coefficients at 140 keV, and the result is used to create attenuation maps at 140 keV from given CT images. This technique has been applied to small pigs in myocardial perfusion studies. Correcting for attenuation increased the image-derived myocardial radionuclide concentration values from 16% of the values observed for excised myocardium to 35% (ignoring background). Adding an additional correction for partial volume effects increases the image-derived value to 93% of the expected value (ignoring background). Precision, however, remains limited.
机译:我们介绍了通过使用从X射线CT数据得出的衰减图来对发射断层扫描进行衰减校正的实验结果。 X射线和伽马射线数据均在包含第三代X射​​线CT扫描仪和单头SPECT扫描仪的原型发射-透射成像系统(“ X-SPECT”)上获得。两个商业成像设备并列放置,以使CT工作台可以在X射线扫描后将患者直接移动到SPECT扫描仪中。这使得能够对SPECT和X射线CT图像进行系统配准。为了校准衰减图的创建,获取了一组针对特定体模几何体和X射线技术的CT扫描,每次扫描中都有不同的衰减材料。将分段线性拟合应用于CT数和140 keV下的已知衰减系数,并将结果用于从给定的CT图像创建140 keV下的衰减图。这项技术已应用于心肌灌注研究中的小猪。衰减的校正将图像来源的心肌放射性核素浓度值从所切除的心肌观察值的16%增加到35%(忽略背景)。添加针对部分体积效果的附加校正,会将图像派生值增加到期望值的93%(忽略背景)。但是,精度仍然有限​​。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号