Count-rate dependent sensitivity in position-sensitive photomultiplier tubes

机译:位置敏感光电倍增管中的计数依赖性灵敏度

获取原文

摘要

A significant count-rate dependence of mean scintillation-pulse amplitude has been observed in prototype small gamma cameras employing NaI(Tl) crystals coupled to Hamamatsu R2487 and R3941 position-sensitive photomultiplier tubes (PSPMTs). The magnitude of the effect is position dependent. Within each camera's 56 mm/spl times/56 mm field-of-view, as the scintillation rate is increased from 500 to 10,000 counts per second, the response to 140 keV gamma emissions from a collimated source of Tc-99m increases by approximately 2-8% in the R2487 and 1-6% in the R3941. The count-rate dependent sensitivities of both tubes were investigated by varying the rate and position of LED pulses directed at the face of the tubes with a 1 mm diameter optical fiber. Within the specified effective area of the R2487, the increase in response from a 500 Hz rate to a 100 kHz rate varied from 1.5% in the center to 20% at one edge; corresponding increases in the R3941 were only 1% and 8%. In both cases, most of the response increase occurred between 500 Hz and 10 kHz. Similar LED tests on conventional photomultiplier tubes resulted in no more than 2% increases.
机译:已经在采用与Hamamatsu R2487和R3941位置敏感的光电倍增管(PSPMT)耦合到Hamamatsu R2487和R3941位置敏感的光电管(PSPMT)的原型小伽马摄像机中观察到平均闪烁脉冲幅度的显着计率依赖性。效果的大小是依赖于位置。在每个相机的56 mm / spl时间/ 56 mm视野下,随着闪烁率从每秒500到10,000计数增加,响应来自TC-99M的准直源的140keV伽马排放量增加了大约2 R3941中R2487中的-8%和1-6%。通过改变具有1mm直径光纤的管子的面部的LED脉冲的速率和位置来研究两个管的计数依赖性敏感性。在R2487的规定有效面积内,从500Hz速率的响应增加到100kHz速率的响应从中心的1.5%变化到一个边缘的20%; R3941的相应增加仅为1%和8%。在这两种情况下,大部分响应都会发生在500Hz和10 kHz之间。在传统的光电倍增管上的类似LED测试导致不超过2%的增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号