Count-rate dependent sensitivity in position-sensitive photomultiplier tubes

机译:位置敏感光电倍增管中与计数率有关的灵敏度

获取原文

摘要

A significant count-rate dependence of mean scintillation-pulse amplitude has been observed in prototype small gamma cameras employing NaI(Tl) crystals coupled to Hamamatsu R2487 and R3941 position-sensitive photomultiplier tubes (PSPMTs). The magnitude of the effect is position dependent. Within each camera's 56 mm/spl times/56 mm field-of-view, as the scintillation rate is increased from 500 to 10,000 counts per second, the response to 140 keV gamma emissions from a collimated source of Tc-99m increases by approximately 2-8% in the R2487 and 1-6% in the R3941. The count-rate dependent sensitivities of both tubes were investigated by varying the rate and position of LED pulses directed at the face of the tubes with a 1 mm diameter optical fiber. Within the specified effective area of the R2487, the increase in response from a 500 Hz rate to a 100 kHz rate varied from 1.5% in the center to 20% at one edge; corresponding increases in the R3941 were only 1% and 8%. In both cases, most of the response increase occurred between 500 Hz and 10 kHz. Similar LED tests on conventional photomultiplier tubes resulted in no more than 2% increases.
机译:在使用耦合到Hamamatsu R2487和R3941位置敏感光电倍增管(PSPMT)的NaI(Tl)晶体的原型小型伽马相机中,已经观察到平均闪烁脉冲幅度的显着计数率依赖性。影响的大小取决于位置。在每台摄像机的56 mm / spl次/ 56 mm视野内,随着闪烁速率从每秒500个增加到10,000个计数,对来自TC-99m准直源的140 keV伽马发射的响应增加了大约2 R2487为-8%,R3941为1-6%。通过改变直径为1 mm的光纤对准管子表面的LED脉冲的速率和位置,研究了两个管子的计数率相关灵敏度。在R2487的指定有效区域内,响应从500 Hz速率增加到100 kHz速率,从中心的1.5%到边缘的20%不等; R3941的相应增加仅为1%和8%。在这两种情况下,大多数响应增加都发生在500 Hz至10 kHz之间。在常规光电倍增管上进行的类似LED测试导致增加幅度不超过2%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号