The Ultraviolet Coronagraph Spectrograph (UVCS) is an optical instrument to be flown on the European spacecraft Solar Heliospheric Observatory (SOHO) to the sun-earth L1 point. The stability requirements of the instrument require the moisture content of the pseudoisotropic composite material not to exceed 0.06% during alignment and calibration. This paper describes the steps necessary to meet this requirement. These steps include a dynamic moisture content analysis, selection of bake out conditions and moisture controls, verification sampling, and use of witness specimens to monitor the moisture content during the prelaunch life of the structure.
机译:紫外线矩形光谱仪(UVCS)是欧洲航天器太阳能幽光素天文台(SOHO)到太阳地球L1点的光学仪器。仪器的稳定性要求要求在对准和校准期间不超过0.06%的伪同步复合材料的水分含量。本文介绍了满足此要求所需的步骤。这些步骤包括动态水分含量分析,选择烘烤条件和水分控制,验证采样和证人试样的使用,以监测结构的预先生寿命期间的水分含量。