Reliability of spectral analysis of the heart rate variability signal from a commercial Holter system

机译:商业龙洞系统心率变异信号的光谱分析可靠性

获取原文

摘要

Power spectral analysis of the heart rate variability signal coming from a traditional Holter recording, might be contaminated by several sources of error. The authors performed a statistical characterization of the measurement error (ME) of some common spectral parameters obtained analyzing the 24-hour R-R intervals list generated by a commercial Holter system. Two ECG signals with known modulation of the R-R interval were generated into a PC and fed to three Holter recorders successively equipped with four different tapes. One ECG signal had 'normal' variability (var=2376 msec/sup 2/), while the other 'reduced' variability (var=86 msec/sup 2/). Twenty-four 10 minute recordings were analyzed on each tape. For each investigated parameter the ME was computed subtracting its true value obtained before Holter recording from its value estimated from the Holter generated R-R list. It was found that the investigated Holter equipment can affect power spectral estimates on the reduced variability tachogram.
机译:来自传统历史记录的心率可变性信号的功率谱分析可能被几个错误源污染。作者对分析由商业HOLTER系统生成的24小时R-R间隔列表的一些常见光谱参数进行测量误差(ME)的统计表征。将具有已知R-R间隔调制的两个ECG信号产生到PC中,并递送到三个孔记录器,连续配备四个不同的胶带。一个ECG信号具有“正常”可变性(VAR = 2376 MSEC / SUP 2 /),而另一个“降​​低”可变性(var = 86毫秒/ sup 2 /)。在每个磁带上分析了二十四分钟录音。对于每个调查的参数,将ME减去了从从HOSTER生成的R-R列表估计的值之前获得的真实值。发现调查的HOLTER设备可以影响降低变化的TACHAGAL对功率谱估计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号