INSTALLATION AND OPERATION ERRORS IN GAS MEASUREMENT

机译:气体测量中的安装和操作误差

获取原文

摘要

Installation errors may occur when an instrument is used in a manner different from how it was calibrated. For example, suppose that a temperature sensor is calibrated in a stirred, constant temperature bath. During calibration, the sensor is in thermal equilibrium with the circulating fluid, and the fluid and sensor temperatures are the same. Next, let the same sensor be used to measure the temperature of natural gas flowing through a pipe at low velocity. If the pipe wall temperature is different from the flowing gas temperature, convection heat transfer will occur between the gas and the pipe wall, conduction heat transfer will occur between the pipe wall and the sensor, and convection heat transfer will occur between the sensor and the flowing gas. In this example, the sensor would not be in thermal equilibrium with the flowing gas and the sensor temperature would be different from the flowing gas temperature.
机译:当以与其校准的方式使用仪器时,可能会发生安装错误。例如,假设在搅拌的恒温浴中校准温度传感器。在校准期间,传感器具有循环流体的热平衡,并且流体和传感器温度是相同的。接下来,使相同的传感器用于测量在低速下流过管道的天然气的温度。如果管壁温度与流动气体温度不同,则气体和管壁之间会发生对流热传递,在管壁和传感器之间发生导电传热,并且在传感器之间发生对流传热发生流动的气体。在该示例中,传感器不会与流动气体的热平衡,传感器温度与流动气体温度不同。