Uncertainty factors in time-interval measurements in ballistocardiography

机译:脉珠上的时间间隔测量中的不确定性因素

获取原文
获取外文期刊封面目录资料

摘要

There is a growing interest on measuring time intervals between the J peak of the ballistocardiogram (BCG) and several features of other cardiovascular signals such as the ECG in order to obtain cardiovascular function markers. Nevertheless, possible uncertainty factors involved in these measurements have not been identified and analyzed, which is a necessary step to advance towards standardization in this reemerging biomedical engineering field. In this paper we analyze the effect of the low-pass cutoff frequency and phase characteristic of filters, noise and power line interference on J-peak time measurement. We conclude that BCG acquisition systems require a minimum low-pass cutoff frequency of 25 Hz but the phase angle of filters at this frequency introduce more than 6 ms delay that increases with increasing filter order. Further, SNR and power line interference levels respectively below about 25 dB and 30 dB, commonly found in data acquisition systems in this area, may lead to uncertainties in the time position of the J peak of tens of milliseconds, which are comparable to measured time interval changes that may have diagnostic interest.
机译:对滚珠通知仪(BCG)的J峰之间的测量时间间隔和其他诸如ECG的其他心血管信号的若干特征的测量时间间隔越来越感兴趣地兴趣,以获得心血管功能标记。然而,尚未确定并分析这些测量中涉及的可能的不确定性因素,这是在该重新制定生物医学工程领域进行标准化的必要步骤。在本文中,我们分析了对滤波器,噪声和电力线干扰对j峰值时间测量的低通截止频率和相位特性的影响。我们得出结论,BCG采集系统需要25Hz的最小低通截止频率,但该频率的滤波器的相位角引入超过6ms的延迟,随着滤波器顺序增加而增加。此外,在该区域的数据采集系统中分别在约25dB和30dB以下的SNR和电力线干扰水平可能导致J峰的数十毫秒的时间位置中的不确定性,这与测量的时间相当可能具有诊断兴趣的间隔更改。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号