首页> 外文会议>Solid Freeform Fabrication annual international Symposium >THE EFFECTS OF DRY HEAT STERILIZATION ON PARTS USING SELECTIVE LASER SINTERING
【24h】

THE EFFECTS OF DRY HEAT STERILIZATION ON PARTS USING SELECTIVE LASER SINTERING

机译:用选择性激光烧结对干热灭菌对零件的影响

获取原文
获取外文期刊封面目录资料

摘要

Selective Laser Sintering (SLS) is a manufacturing process that can build arbitrarily shaped parts without part specific tooling. Its advantages have been employed in many different fields, one of these being medical surgery. Currently, SLS is limited in medical applications as a pre-operative modeling tool. For SLS manufacturing to progress in areas like compliant surgical tooling and patient specific bone matrices, concurrent work is needed to investigate the effects of medical sterilization on SLS materials. This paper presents the results of sterilization experiments on SLS parts built from nylon 11. To simulate the process of introducing tools into a sterile environment, these specimens were subjected to multiple rounds of dry heat sterilization. Changes to the dimensions, tensile strength and flexibility were recorded and analyzed. It was found that the specimens' dimensions remained relatively constant. Both the tensile modulus and the flexural modulus decreased as the sterilization cycles progressed. The tensile modulus decreased by 25% and the flexure modulus decreased by 19% after ten rounds of sterilization.
机译:选择性激光烧结(SLS)是一种制造工艺,可以在没有部件特定的工具的情况下构建任意形状的零件。它的优点是许多不同的领域,其中一个是医学手术。目前,SLS在医学应用中受到限制作为预惯用的建模工具。对于SLS制造业,在符合符合的外科手术和患者特异性骨基质等地区的进展中,需要并行工作来研究医疗灭菌对SLS材料的影响。本文介绍了由尼龙11建造的SLS部件的灭菌试验结果。为了模拟将工具引入无菌环境的过程,这些样品受到多轮干热灭菌。记录和分析尺寸,拉伸强度和柔韧性的变化。发现标本的尺寸保持相对恒定。随着灭菌循环的进展,拉伸模量和弯曲模量均降低。抗拉模量降低25%,弯曲模量在十轮灭菌后减少了19%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号