首页> 外文会议>IEEE Nuclear Science Symposium Conference Record >New calibration and evaluation method for PET scanners using point-like radioactive sources
【24h】

New calibration and evaluation method for PET scanners using point-like radioactive sources

机译:使用点状放射源的PET扫描仪的新校准和评估方法

获取原文

摘要

Objective: In order to improve the convenience and reliability of the calibration procedure with respect to the quantitative aspects of PET scanners, we have been developing new methods that rely on the use of point-like radioactive sources. The aim of this study is to report the results of the application of the proposed method for obtaining the cross-calibration factors (CCFs) of a clinical PET scanner for the first time. Methods: A ~(22)Na point-like source was moved in the axial direction to cover the axial field-of-view (FOV) of a commercial PET scanner, SET-2400W (Shimadzu). From reconstructed images, saturated regions-of-interest (ROI) values were used to calculate the CCFs that were defined slice by slice. Results: It was possible to obtain CCFs that were practically equivalent to those obtained by the standard method with cylindrical water phantoms. The variations in the CCFs obtained over about a half-year were well reproduced by the proposed method. Conclusion: The proposed calibration method, which is based on the use of a point-like radioactive source, is useful for determining CCFs and their variations for a clinical PET scanner.
机译:目的:为了提高宠物扫描仪的定量方面的校准程序的便利性和可靠性,我们一直在开发依赖于使用点状放射源的新方法。本研究的目的是报告第一次获得临床PET扫描仪的跨校准因子(CCF)的提出方法的应用结果。方法:在轴向方向上移动〜(22)族点状源,以覆盖商业PET扫描仪的轴向视野(FOV),设置-2400W(Shimadzu)。从重建的图像中,使用饱和区域(ROI)值来计算通过切片来计算已定义的切片的CCF。结果:可以获得几乎相当于通过圆柱形水法测量的标准方法获得的CCF。在大约半年中获得的CCF的变化通过该方法再现良好。结论:基于点形放射源的校准方法,可用于确定CCF及其对临床PET扫描仪的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号