首页> 外文会议>Latin American Symposium on Nuclear Physics and Applications >Characterization of HPGe Gamma Spectrometric Detectors Systems for Instrumental Neutron Activation Analysis (INAA) at the Colombian Geological Survey
【24h】

Characterization of HPGe Gamma Spectrometric Detectors Systems for Instrumental Neutron Activation Analysis (INAA) at the Colombian Geological Survey

机译:哥伦比亚地质调查仪器中子激活分析仪器中子激活分析探测系统的特征

获取原文

摘要

This paper presents the progress made by the Neutron Activation Analysis (NAA) laboratory at the Colombian Geological Survey (SGC in its Spanish acronym), towards the characterization of its gamma spectrometric systems for Instrumental Neutron Activation Analysis (INAA), with the aim of introducing corrections to the measurements by variations in sample geometry. Characterization includes the empirical determination of the interaction point of gamma radiation inside the Germanium crystal, through the application of a linear model and the use of a fast Monte Carlo N-Particle (MCNP) software to estimate correction factors for differences in counting efficiency that arise from variations in sample density between samples and standards.
机译:本文介绍了哥伦比亚地质调查(西班牙语首字母缩略词的SGC)中子激活分析(NAA)实验室的进展,旨在引入仪器中子激活分析(INAA)的伽马光谱系统的表征通过样本几何变化对测量的校正。表征包括锗晶体内γ辐射的相互作用点的实证测定,通过应用线性模型和使用快速蒙特卡罗n粒子(MCNP)软件来估计校正因子,以估算计数效率的差异从样品和标准之间的样本密度的变化。