首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Balancing sensitivity gains and method simplification:Investigating HR/AM and a new nanospray source for a challenging bioanalytical method
【24h】

Balancing sensitivity gains and method simplification:Investigating HR/AM and a new nanospray source for a challenging bioanalytical method

机译:平衡敏感性收益和方法简化:研究人力资源/ am和新的纳米浦背源,以挑战生物分析方法

获取原文
获取外文期刊封面目录资料

摘要

Evaluated four different methodologies using combination of HRAM MS, HILIC, UHPLC, Nanospray/nano flow. Method 1 (HILIC, isocratic): (1)Direct transfer to 1D HILIC chromatography was not successful; Method 2 (HILIC, forward flush): (2) Acceptable result but too sensitive to aqueous content. Method 3 (Phenyl Hexyl, UHPLC): (1) Most promising. No carryover. PLs under control with backflush; Method 4 (C18, Nanocolumn and Easy spray source): (2) Good precision and accuracy. But can't do column backflush. Handling of phospholipids are still a challenge.
机译:使用HRAM MS,HILIC,UHPLC,纳米涂布/纳米流量的组合评估四种不同的方法。方法1(HILIC,BIRACIC):(1)直接转移至1D HILIC色谱法不成功;方法2(HILIC,正向冲洗):(2)可接受的结果,但对含水含量过于敏感。方法3(苯基己基,UHPLC):(1)最有前途。没有残留。通过反冲的控制下来;方法4(C18,纳米,易于喷涂源):(2)良好的精度和精度。但不能做柱子反弹。磷脂的处理仍然是一项挑战。