首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, and Improved Robustness of Data
【24h】

Second-Generation Electron Transfer Dissociation (ETD) on the Thermo Scientific Orbitrap Fusion Mass Spectrometer with Improved Functionality, Increased Speed, and Improved Robustness of Data

机译:第二代电子传递解离(ETD)在Thermo Scientific orbiTrap融合质谱仪上,具有改进的功能,速度提高和数据的鲁棒性

获取原文

摘要

A second-generation glow-discharge-based Easy-ETD source has been developed for the Orbitrap Fusion MS that incorporates significant hardware and software advancements. Calibration of the reaction kinetics removes the guesswork from ETD, and leads to conditions that optimizes the ETD reaction and scan cycle time. The calibrated reaction conditions are demonstrated to provide optimal conditions for ETD identifications and sequence coverage. Parallel acquisition provides ITMS~2 ETD scan rate cycle times of up to 12 Hz.
机译:已经开发了一种基于第二代发光放电的易于ETD源,用于融合具有重要硬件和软件进步的orbitrap融合MS。反应动力学的校准去除ETD的猜测,并导致优化ETD反应和扫描循环时间的条件。证明校准的反应条件为ETD识别和序列覆盖提供最佳条件。并行获取提供了最多12 Hz的ITMS〜2 ETD扫描速率周期时间。