首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Improving Quantification in Time of Flight (TOF) Mass Spectrometers by Optimizing Detector Response
【24h】

Improving Quantification in Time of Flight (TOF) Mass Spectrometers by Optimizing Detector Response

机译:通过优化探测器响应,改善飞行时间(TOF)质谱仪的量化

获取原文

摘要

The electron emission process that all electron multipliers rely on is a random process that is typically described with Poisson statistics. The output of electron multipliers is based on repeated electron emission events that occur on many discrete dynodes or along a continuous dynode structure. Calculating the pulse height distribution (the probability of measuring a pulse of a particular height for a single input event) from secondary electron yield data is difficult, but properties of pulse height have been calculated and are summarized by the curves shown in Figure 1.
机译:所有电子乘法器依赖的电子发射过程是通常用泊松统计描述的随机过程。电子乘法器的输出基于在许多离散Dynodes或沿连续的Dynode结构上发生的重复的电子发射事件。计算来自二级电子屈服数据的脉冲高度分布(测量单个输入事件的特定高度脉冲的概率)是困难的,但已经计算了脉冲高度的性质,并通过图1中所示的曲线来概括。