Isolating the Effects of MicroChannel Plate Parameters on the Ultimate Resolution in Time-Of-Flight Mass Spectrometers

机译:隔离微通道板参数对飞行时间质谱仪的最终分辨率的影响

获取原文
获取外文期刊封面目录资料

摘要

TOF mass resolution can be improved with the use of MCPs that utilize smaller pore diameters and steeper bias angles with no measured loss in detection efficiency. Changing from a 5 (mu)m diameter pore / 12 deg bias angle MCP to a 2 (mu)m diameter pore/19 deg bias angle MCP improved the mass resolution by approx15percent for masses of 1000-3000 Da. The flatness of a mounted MCP is determined by the hardware and the design of the detector. Proper designs can routinely produce detectors with input surface flatness values <15 (mu)m over 25mm diameter areas. Improperly designed or assembled hardware can distort otherwise flat MCP.
机译:通过使用使用较小的孔径和较陡峭的偏置角度,可以改善TOF质量分辨率,并且在检测效率上没有测量损失。从5(mu)的直径孔/ 12°偏置角MCP转换为2(mu)M直径孔/ 19°偏置角MCP将质量分辨率提高了大约150-3000Da的质量。安装的MCP的平坦度由硬件和检测器的设计决定。适当的设计可以经常生产具有输入表面平坦度值<15(MU)M超过25mm的区域的探测器。设计不当或组装的硬件可以扭曲其他平面MCP。