首页> 外文会议>International Solid-State Sensors, Actuators and Microsystems Conference >PPB-LEVEL DETECTION OF BENZENE DILUTED IN WATER BY BUBBLING EXTRACTION SYSTEM AND UV SPECTROSCOPY BASED MEASUREMENTS
【24h】

PPB-LEVEL DETECTION OF BENZENE DILUTED IN WATER BY BUBBLING EXTRACTION SYSTEM AND UV SPECTROSCOPY BASED MEASUREMENTS

机译:通过鼓泡萃取系统和UV光谱法测量在水中稀释的苯的PPB水平检测

获取原文
获取外文期刊封面目录资料

摘要

A new and easy-to-use portable sensor dedicated to the detection of benzene diluted in aqueous solution is presented. By connecting a bubbling device to a high efficiency UV detection cell, previously developed for a BTX air monitoring system, we could achieve a detection limit down to 100 ppb with a S/N of about 3. This sensitivity conforms to wastewater regulations in both America and Japan (resp. 500 and 120 ppb). The sensor also exhibits a linear response within the 0.1 - 3 ppm range and reproducibility. Measurements take less than 10 minutes, which makes this sensor particularly suitable for fast and accurate on-field monitoring applications.
机译:提出了一种新的和易于使用的便携式传感器,其专用于在水溶液中稀释的苯的检测。通过将鼓泡装置连接到高效的UV检测单元,以前为BTX空气监测系统开发,我们可以通过约3的S / N实现检测限直到100ppb。这种敏感性符合美国的废水法规和日本(resp。500和120 ppb)。传感器还在0.1-3ppm范围内具有线性响应和再现性。测量需要不到10分钟,这使得该传感器特别适用于快速准确的现场监测应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号