首页> 外文会议>International Workshop on Low Temperature Detectors >Performance of an X-ray single pixel TES microcalorimeterunder DC and AC biasing
【24h】

Performance of an X-ray single pixel TES microcalorimeterunder DC and AC biasing

机译:X射线单像素TES微量电流仪DC和AC偏置的性能

获取原文

摘要

We are developing Frequency Domain Multiplexing (FDM) for the read-out of TES imaging microcalorimeter arrays for future X-ray missions like IXO. In the FDM configuration the TES is AC voltage biased at a well defined frequencies (between 0.3 to 10MHz) and acts as an AM modulating element. In this paper we will present a full comparison of the performance of a TES microcalorimeter under DC bias and AC bias at a frequency of 370kHz. In both cases we measured the current-to-voltage characteristics, the complex impedance, the noise, the X-ray responsivity, and energy resolution. The behaviour is very similar in both cases, but deviations in performances are observed for detector working points low in the superconducting transition (R/R_N < 0.5). The measured energy resolution at 5.89keV is 2.7eV for DC bias and 3.7eV for AC bias, while the baseline resolution is 2.8eV and 3.3eV, respectively.
机译:我们正在开发用于读出的TES成像微量电流表阵列的频域多路复用(FDM),以便将未来的X射线任务如IXO。在FDM配置中,TES是在良好定义的频率(0.3至10MHz)处偏置的交流电压,并用作AM调制元件。本文我们将在370kHz的频率下完全比较TES微量微量电磁表的性能和AC偏置。在这两种情况下,我们测量了电流到电压特性,复杂阻抗,噪声,X射线响应度和能量分辨率。这种情况在这两种情况下都非常相似,但是在超导转变中的检测器工作点(R / R_N <0.5)低,观察到性能的偏差。 5.89Kev的测量能量分辨率为2.7EV,用于DC偏置和AC偏置的3.7EV,而基线分辨率分别为2.8EV和3.3EV。