There are many commercial WBV platforms for clinical applications but in literature very few have been tested to verify their actual amplitude, frequency spectra and waveform shape of vibration, although some clinical studies showed that these parameters can influence significantly the muscles performance. In this study the performances of a prototype WBV_1 designed by the authors are compared with a commercial platform WBV_2 typically used in clinical and rehabilitation investigations. Tests performed with four accelerometers in 3 different directions show the presence of transversal accelerations that in some cases exceeded the vertical one in WBV_2 while they were always lower than the 25% in WBV_1. Also the waveform shape of the vibration provided by WBV_1, evaluated by means of the SINAD parameter, show a better value compared to the commercial device: indeed secondary harmonics reaches the 30% of the fundamental in WBV_2 but they are always below the 25% in the prototype.
机译:有许多用于临床应用的商业WBV平台,但在文献中,很少已经测试过验证其实际幅度,频谱和波形的振动形状,尽管一些临床研究表明这些参数可以显着影响肌肉性能。在本研究中,作者设计的原型WBV_1的性能与通常用于临床和康复调查的商业平台WBV_2进行比较。在3个不同方向上用四个加速度计进行的测试显示出横向加速的存在,在某些情况下,在某些情况下超过WBV_2中的垂直一个,而它们总是低于WBV_1中的25%。此外,WBV_1提供的振动的波形形状,通过SINAD参数评估,与商业设备相比,更好的价值:实际上二次谐波达到WBV_2中基本的30%,但它们总是低于25%原型。