首页> 外文会议>International Mass Spectrometry Conference >STUDY FOR MASS RESOLUTION OF QUADRUPOLE MASS SPECTROMETERS ACCORDING TO THE ROD ELECTRODE ALIGNMENT
【24h】

STUDY FOR MASS RESOLUTION OF QUADRUPOLE MASS SPECTROMETERS ACCORDING TO THE ROD ELECTRODE ALIGNMENT

机译:根据杆电极对准的四极孔质谱仪质量分辨率研究

获取原文

摘要

Radio frequency (RF) quadrupole mass spectrometers (QMS) have come to be used in various fields because of their compactness. Ionized specimens are separated while passing through a QMS in accordance with their m/z values. The mass resolution of QMSs must be raised to identify species with high accuracy. The mass resolution is related to the alignment of a QMS's four rod electrodes [1].
机译:由于其紧凑性,射频(RF)四极谱质量光谱仪(QMS)已经在各个领域中使用。根据其M / Z值通过QMS分离电离样品。必须提高QMS的质量分辨率以识别高精度的物种。质量分辨率与QMS的四个杆电极的对准有关[1]。