COMPARISON OF CCS(N_2) MEASUREMENTS OBTAINED FROM TWO DIFFERENT T-WAVE ION MOBILITY SYSTEMS WITH DIRECT MEASUREMENTS USING A DRIFT TUBE ION MOBILITY SYSTEM

机译:使用漂移管离子移动系统的两种不同T波离子移动系统获得的CCS(N_2)测量的比较

获取原文
获取外文期刊封面目录资料

摘要

Excellent correlation has been observed between in house and literature drift tube CCS(N_2) values. It has been demonstrated that the CCS(N_2) values determined by calibrated T-wave systems (of either geometry) are equivalent to those obtained by a linear field drift tube IM systems. The term 'deviation' should be used to describe observed variation between experimentally measured CCS(N_2) values not 'error' or 'accuracy' as these terms imply that a given experimentally observed value is correct and free from measurement error with respect to another experimentally observed value.
机译:在房屋和文献漂移管CCS(N_2)值之间观察到出色的相关性。已经证明,由校准的T波系统(几何形状)确定的CCS(N_2)值等同于通过线性场漂移管IM系统获得的值。术语“偏差”应用于描述实验测量的CCS(N_2)值之间观察到的变化,因为这些术语意味着给定的实验观察值是正确的,并且没有关于另一个实验相对于另一个的测量误差的误差观察到的价值。