Single Field Direct Drift Time to CCS Calibration for a Linear Drift Tube Ion Mobility Mass Spectrometer

机译:单场直接漂移时间为CCS校准线性漂移管离子迁移率质谱仪

获取原文

摘要

A novel high speed technique is developed for calibration of the CCS domain for a linear drift IM-MS spectrometer. The technique allows fast and routine CCS measurements in a single LC run with pre/post or in-run calibration. Current technique requires to have info on both m/z and z of an ion to identify its CCS based on the CCS of two known ions and therefore it is perfectly matching IM to MS tandem instrumentation.
机译:开发了一种新的高速技术,用于校准线性漂移IM-MS光谱仪的CCS域。该技术允许在单个LC中进行快速和常规CCS测量,使用预/柱或运行校准运行。目前的技术需要在离子的M / Z和Z上具有信息,以基于两个已知离子的CC来识别其CCS,因此它与MS Tandem仪器完全匹配。