Quantitative and qualitative analysis using ASAP ionization combined with SPME sample introduction on an Orbitrap mass spectrometer.

机译:用ASAP电离与SPME样品介绍的定量和定性分析在露侧谱仪上介绍。

获取原文

摘要

SPME is easily adaptable to ASAP methodology. higher sheath gas flow rate does not significantly affect either peak area or peak reproducibility, but impacts the noise level. the APCI probe temperature can be programmed to effectively desorb the analytes from SPME fibers. feasibility of quantitative operation of SPME ASAP was demonstrated by obtaining good quality calibration curves for both types of SPME fibers. the limit of detection (LOD) for majority of investigated analytes was < 0.5 ng/mL with SPME ASAP. The same analytes run just on an ASAP probe gave LODs approximately 0.5 (mu)g/mL (1000x higher).
机译:SPME很容易适应ASAP方法。较高的鞘气流流速不会显着影响峰面积或峰值再现性,但影响噪声水平。 APCI探针温度可以被编程为有效地解吸来自SPME纤维的分析物。通过获得两种类型的SPME纤维的良好质量校准曲线,证明了SPME的定量操作的可行性。大多数研究分析物的检测极限(LOD)为<0.5ng / ml,SPME尽快。在ASAP探针上运行的相同的分析物给出了大约0.5(mU)g / ml(1000倍更高的)。