The benefit of pulsatile perfusion during cardiopulmonary bypass is still controversial. In this study we examined the non-pulsatile perfusion generated by a roller pump, the pulsatile perfusion induced by a roller-pump and the "physiological" pulsatile perfusion generated by a new device on renal and gut tissue perfusion, haemolysis and inflammatory response. In the group with the new computer-controlled pump system there was no significant difference for haemodynamic data between the measurement before and under ECC. The degree of haemolysis was similar in both pulsatile groups but less in the non-pulsatile group. There was no difference on serum concentration of IL 6, IL 1ra and on regional blood flow between the groups. There are no advantages for pulsatile perfusion on haemolysis, inflammatory response and regional organ blood flow.
机译:心肺旁路期间脉动灌注的益处仍然存在争议。在这项研究中,我们检查了由滚筒泵产生的非脉动灌注,由辊泵引起的脉动灌注和由新装置产生的肾脏和肠道组织灌注,溶血和炎症反应产生的“生理学”脉动灌注。在新的计算机控制泵系统中,ECC之前和下的测量之间的血液动力学数据没有显着差异。在脉腭基团中,溶血基团在非脉粘剂组中较少。 IL 6,IL 1RA的血清浓度和组之间的区域血流没有差异。对溶血性灌注对溶血,炎症反应和区域器官血流没有任何优点。