首页> 外文会议>International Conference on Radionuclide Metrology and its Applications >Monte Carlo calculations of calibration and geometry correction factors of a radionuclide calibrator
【24h】

Monte Carlo calculations of calibration and geometry correction factors of a radionuclide calibrator

机译:放射性核素校准器校准和几何校正因子的蒙特卡罗计算

获取原文

摘要

The activity of radioactive pharmaceuticals administered to patients in nuclear medicine is usually determined using well-type high-pressureionization chambers. For the Bqmeter chamber (Consortium BQM, Czech Republic) a Monte Carlo model was created using the MCNP4C2 code. Basic chamber characteristics for two sample containers of various geometry (a vial and an ampoule) were calculated and compared with measurements. As the pharmaceuticals are often measured in various syringes, the chamber response for samples in syringes was also studied.
机译:施用给核医学患者的放射性药物的活性通常使用良好型高压缩室确定。对于BQMETER分庭(Consortium BQM,捷克共和国)使用MCNP4C2代码创建了Monte Carlo模型。计算并与测量相比,计算两种样品容器的基本腔室特性(小瓶和安瓿)。由于药物通常在各种注射器中测量,因此还研究了注射器中样品的腔室响应。