首页> 外文会议>International Conference on Radionuclide Metrology and its Applications >Optimization of counting times for short-lived gamma-ray emitters in air filter samples
【24h】

Optimization of counting times for short-lived gamma-ray emitters in air filter samples

机译:空气过滤器样品中短缘伽马射线发射器的计数时间优化

获取原文

摘要

A methodology for the optimization of the counting times in a series of measurements of gamma-ray emitters in air filters is presented.In the optimal measurement regime in measurements of all the filters in a batch, the same minimum detectable activity is attained. It is shown how the number of filters, the properties of the gamma-ray emitter and the equipment influence the measurement time of the batch of filters and the minimum detectable activity attained.
机译:提出了一种用于优化空气过滤器中的伽马射线发射器的一系列测量的计数时间的方法。在批次中的所有滤波器的测量中的最佳测量方案中,实现了相同的最小可检测活动。示出了如何滤波器的数量,伽马射线发射器的性质和设备影响批量过滤器的测量时间和最低可检测活动。