首页> 外文会议>International Workshop on Digital Mammography(IWDM 2006) >Use of the European Protocol to Optimise a Digital Mammography System
【24h】

Use of the European Protocol to Optimise a Digital Mammography System

机译:使用欧洲协议来优化数字乳腺X线摄影系统

获取原文

摘要

An experimental method of determining the optimal beam qualities and doses for digital mammography systems is described, and applied to a CR system. The mean glandular dose (MGD) and contrast-to-noise ratio (CNR) were measured using phantoms. For each thickness of phantom a range of kV and target/filter combinations were tested. Optimal beam quality was defined as that giving a target CNR for the lowest MGD. The target CNR was that necessary to achieve at least the minimum standard of image quality defined in European Guidance. An inverse relationship between CNR and threshold contrast was confirmed over a range of thicknesses of PMMA and different beam qualities and doses. Optimisation indicated that relatively high energy beam qualities (e.g. 31 kV Rh/Rh) should be used with a greater detector dose to compensate for the lower contrast when compared to using lower energy X-rays. The results also indicate that current AEC designs that aim for a fixed detector dose are not optimal.
机译:描述了一种确定数字乳腺X线摄影系统的最佳光束素质和剂量的实验方法,并应用于CR系统。使用幽灵测量平均腺剂量(MGD)和对比度 - 噪声比(CNR)。对于每个厚度的幽灵,测试了一系列KV和目标/滤波器组合。最佳光束质量被定义为为最低MGD提供目标CNR。目标CNR是必须至少在欧洲指导中定义的最小图像质量标准。在PMMA的一系列厚度和不同的光束品质和剂量之间证实了CNR与阈值对比之间的反比关系。优化表明,与使用较低能量X射线相比,应与更高的检测器剂量一起使用相对高的能量光束质量(例如31kV rh / Rh)以补偿较低的对比度。结果还表明,目前AEC设计的目前的瞄准固定探测器剂量是不是最佳的。