首页> 外文会议>International Flow Measurement Conference >Performance Evaluation of Critical Flow Venturi Nozzle with Primary Standard Calibration Facilities at FCRI and Validation with ISO Standards
【24h】

Performance Evaluation of Critical Flow Venturi Nozzle with Primary Standard Calibration Facilities at FCRI and Validation with ISO Standards

机译:FCRI主要标准校准设施临界流动术的绩效评估与ISO标准的验证

获取原文

摘要

A 2" NB Critical Flow Venturi Nozzle (CFVN) of Nominal Flow capacity 25m~(3)/h was designed at Fluid Control Research Institute (FCRI), India as a transfer standard for undertaking calibration of smaller capacity flow meters. The Nozzle fabricated was of Toroidal throat design and was assembled as per the general guidelines of ISO 9300:2005 "Measurement of gas flow by means of Critical flow Venturi nozzles". The Air Flow Measurement facilities at FCRI is equipped with Primary Standards Pressure-Volume-Temperature-time (PVTt) facility of 2m3 Nominal volume and 90 m~(3)/h maximum flow capacity and a Bell Prover of 500 ltr volume and with a maximum flow capacity of 40m~(3)/h. The performance of the Critical Flow Venturi Nozzle assembly was tested and analyzed with both PVTt facility and Bell Prover. PVTt facility can be used as a collection tank for estimation of mass flow rate during operation of CFVN during suction mode. To establish dynamic traceability in measurements and compare the test results with other proven primary standards, the same CFVN assembly was calibrated both under pressure mode and suction mode using the Primary standard Bell Prover system available at the Primary Standard Air Flow Laboratory at FCRI. During filling mode operation of Bell Prover, the Throat Reynolds number of nozzle was about 1.24E+05 and in Discharge mode of flow test the Throat Reynolds number of nozzle was about 8.61E+04. Cd of the nozzle was established at different Reynolds numbers with both PVTt facility and Bell prover. The results obtained with the above methods were validated using ISO equation given in ISO 9300: 2005.
机译:在流体控制研究所(FCRI)中设计了一个2“Nb临界流动文丘里喷嘴(CFVN)的标称流量容量25m〜(3)/ h,作为开展较小容量流量计的校准的转移标准。喷嘴制造是环形喉部设计,并根据ISO 9300:2005的一般指南组装,通过临界流动文丘里喷嘴测量气体流动“。FCRI的空气流量测量设施配备了主要标准压力量 - 温度 - 时间(PVTT)设施为2M3标称体积,90米〜(3)/ H最大流量和500 LTR卷的钟礼品,最大流量为40m〜(3)/ h。临界流的性能使用PVTT设施和贝尔证报进行了测试和分析了文丘里喷嘴组件。PVTT设施可用作收集罐,用于在吸入模式下估计CFVN的速度过程中的质量流量。在测量中建立动态可追溯性并比较测试结果采用其他经过验证的主要标准,使用在FCRI的主要标准空气流实验室的主要标准贝尔证据系统,在压力模式和吸力模式下校准相同的CFVN组件。在烧结模式操作期间,喉部雷诺的喷嘴数约为1.24e + 05,并且在流量测试的排出模式下,喷嘴的喉部雷诺数为约8.61e + 04。喷嘴的CD与PVTT设施和贝尔先驱的不同雷诺数建立。使用ISO 9300:2005中给出的ISO方程验证了用上述方法获得的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号