Validation of Landsat 7 ETM+ band 6 radiometric performance

机译:验证Landsat 7 ETM +带6辐射性性能

获取原文

摘要

Since shortly after launch the radiometric performance of band 6 of the ETM+ instrument on Landsat 7 has been evaluated using vicarious calibration techniques for both land and water targets. This evaluation indicates the radiometric performance of band 6 has been both highly stable and accurate. Over a range corresponding to a factor of two in radiance (5 to 55 C in kinetic temperature terms) the difference between the in-situ derived radiance and the image derived radiance is on average 0.5% or less. Water targets are the easiest to use but are limited to the temperature range from 0 to about 32 C. Land targets can reach 55 C or more but are far less spatially homogeneous than water targets with respect to both local surface temperature and spectral emissivity. The techniques used and the results are described.
机译:由于在推出陆地和水目标的替代校准技术,因此评估了Landsat 7上的ETM +仪器的带6的辐射性能。该评估表明频带6的辐射性能既高度稳定又准确。在对应于辐射倍数的范围内(动力学温度术语5至55℃的5至55℃),原位导出的辐射与图像导出的辐射之间的差异平均为0.5%或更低。水目标是最容易使用的,但仅限于0至约32℃的温度范围内的温度范围,但是土地靶标可以达到55℃或更大但相对于局部表面温度和光谱发射率的水目标远不太均匀。描述了使用的技术和结果。