Signal heterogeneity in contrast-enhanced echocardiographic imaging despite uniform concentration of the contrast agent

机译:在对比度增强超声心动图成像中的信号异质性尽管造成造影剂的均匀浓度

获取原文
获取外文期刊封面目录资料

摘要

To quantify the heterogeneity of contrast enhancement inside left ventricular (LV) cavity, 10 patients were studied by harmonic intermittent power Doppler following venous echo contrast administration. Regions of interest were drawn in the L V cavity along the main axis and transversally to derive time-intensity curves. Signal intensity at plateau (A) significantly decreased along LV longitudinal axis: from apex (48±6 dB), to center (41±7) and base (31±9 dB, p < 0.05), while intensity rise (b) was unaltered. A also was lower in proximity of the septum than in the center of LV cavity (p<0.05). Finally, beat-to-beat variability increased with increasing depth. Thus, L V contrast intensity varies despite a uniform concentration of the contrast agent at each time, due to ultrasound attenuation and heterogeneous beam intensity. Parameter b appears more consistent than A in deriving quantitative flow information.
机译:为了量化左心室(LV)腔内对比增强的异质性,通过静脉回波对比度给药后谐波间歇性电力多普勒研究了10名患者。利益区域沿着主轴横向绘制在L V腔中,横向地推导出时间强度曲线。高原(a)的信号强度沿LV纵向轴线显着降低:从顶点(48±6 dB),中心(41±7)和碱(31±9 dB,P <0.05),而强度上升(b)是unalter。在隔膜的附近也比在LV腔的中心(P <0.05)的邻近较低。最后,随着深度的增加,节拍变异增加。因此,尽管在每次均匀浓度和非均相光束强度,但是由于超声衰减和异质光束强度,因此,尽管造成造影剂的浓度均匀浓度,因此,L V对比度强度变化。参数B看起来比推导定量流信息更致一致。