Comparison of two methods for removing baseline wander in the ECG

机译:两种方法在心电图中删除基线漫步的两种方法

获取原文

摘要

The authors compare the cubic-spline method with multipole null-phase digital filters in their ability to correct for baseline wander on 69 ECG segments with both normal and abnormal rhythms. A single-pole 0.05-Hz filter was included in this study for comparison. A null-phase, 6-pole filter with a cutoff between 0.75 and 1.0 Hz can attenuate low-frequency noise (i.e. correct baseline) as well as the cubic spline. However, such a filter introduces minor changes in the original signal, especially in the presence of a long RR interval. The cubic spline is very dependent on the accurate determination of QRS onset. However, both perform better than the single-pole 0.05-Hz filter, which does very little to attenuate low-frequency noise.
机译:作者将立方样条方法与多极值数字滤波器进行比较,以便在具有正常和异常节律的69个ECG段上纠正基线漫步的能力。本研究中包含单极0.05 Hz过滤器以进行比较。具有0.75和1.0Hz之间的截止值的空相,6极过滤器可以衰减低频噪声(即,正确的基线)以及立方样条。然而,这种过滤器引入了原始信号的次要变化,尤其是在长RR间隔存在下。立方样条依赖性对QRS发作的精确测定非常好。但是,两者都比单极0.05 Hz过滤器更好,这几乎没有衰减低频噪声。