首页> 外文会议>International Symposium on Temperature and Thermal Measurements in Industry and Science >Calibration of an Absolute Radiation Thermometer for Accurate Determination of Fixed-Point Temperatures
【24h】

Calibration of an Absolute Radiation Thermometer for Accurate Determination of Fixed-Point Temperatures

机译:绝对辐射温度计的校准,用于精确测定定点温度

获取原文

摘要

For accurate determinations of thermodynamic temperature, NPL has developed its absolute radiation thermometer (ART), which is calibrated traceably against a cryogenic radiometer. This article reviews some of the potential sources of systematic uncertainty present in the calibration and use of ART. In particular, this article is concerned with the evaluation of the size-of-source effect and the lens transmittance, as well as potential differences in the responsivity of a transfer trap detector when calibrated in terms of radiant power and used in irradiance mode.
机译:为了精确测定热力学温度,NPL开发了其绝对辐射温度计(ART),其可追溯到低温计可追溯校准。本文审查了校准和使用艺术中的系统不确定性的一些潜在来源。特别地,本文涉及评估源极效应和透镜透射率,以及在辐射功率方面校准并在辐照模式下使用时转移陷阱检测器的响应度的潜在差异。