The purpose of this study was to determine if the removal of the periosteum or the application of a strain gage has any significant effect on the structural response of human ribs. A total of 32 three-point bending tests were performed on 16 matched whole rib sections obtained from the left and right sides of five male human thoraces. For one test group, matched specimens were tested to determine the effect of removing the soft tissue and periosteum versus leaving it intact. For a second test group, matched specimens were tested to determine the effects of placing a strain gage on the tension side of the specimen versus no strain gage attachment. The specimens were tested using a servo-hydraulic material testing machine (MTS) at a displacement rate of 17.78 cm/s with a fixed testing span of 10.16 cm. Prior to testing, a microCT was used to obtain a detailed cross-sectional image of each specimen at the point of the impactor blade contact. There were no statistical differences in area moment of inertia (p=0.60), distance to the neutral axis (p=0.29), peak moment (p=0.14), peak impactor displacement (p=0.13), estimated peak stress (p=0.42), or estimated peak strain (p=0.15) between specimens with the periosteum and those without the periosteum. There were no statistical differences in area moment of inertia (p=0.76), distance to the neutral axis (p=0.20), peak moment (p=0.81), peak impactor displacement (p=0.91), estimated peak stress (p=0.59), or estimated peak strain (p=0.29) between specimens with a strain gage and those without a strain gage. In summary, neither the removal of the periosteum nor the application of a strain gage has any significant effect on the structural response of human ribs in dynamic three-point bending.
机译:本研究的目的是确定骨膜的去除还是应变计的施用对人肋的结构响应有任何显着影响。在从五个雄性人父族的左侧和右侧获得的16个匹配的整个肋骨部分进行总共32个三点弯曲试验。对于一个测试组,测试匹配的标本以确定去除软组织和骨膜的效果与完好无损。对于第二次测试组,测试匹配的样品以确定将应变计放在样品的张力侧的效果与菌株计附着。使用伺服液压材料试验机(MTS)以17.78cm / s的位移率测试标本,固定测试距离为10.16厘米。在测试之前,用于在撞击刀片接触点处获得每个样本的详细横截面图像。惯性面积(p = 0.60)没有统计差异,到中性轴的距离(p = 0.29),峰值力矩(p = 0.14),峰值撞击器位移(p = 0.13),估计峰值应力(p = 0.42),或估计在具有骨膜的标本和没有骨膜的那些之间的估计峰菌株(p = 0.15)。面对惯性面积(p = 0.76),距离中性轴(p = 0.20),峰值撞击器位移(p = 0.91),估计峰值应力(p = 0.59),或估计具有应变计的标本与没有应变计的试样之间的峰菌株(p = 0.29)。发明内容既不除去骨膜,也不施加应变计对动态三点弯曲中人肋的结构响应有任何显着影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号