首页> 外文会议>International conference on miniaturized chemical and biochemical analysis systems >INTEGRATED AMORPHOUS SILICON PHOTODIODE DETECTOR FOR MICROFABRICATED CAPILLARY ELECTROPHORESIS DEVICES
【24h】

INTEGRATED AMORPHOUS SILICON PHOTODIODE DETECTOR FOR MICROFABRICATED CAPILLARY ELECTROPHORESIS DEVICES

机译:用于微型制衡毛细管电泳装置的集成非晶硅光电二极管探测器

获取原文

摘要

We have demonstrated that hydrogenated amorphous Si (a-Si:H) photodiodes are sufficiently sensitive to be used as detectors for chemical and biological assays performed on microfabricated capillary electrophoresis devices. A limit of detection of 56 pM for fluorescein flowing in a 50-μm deep microchannel is observed utilizing confocal optics. Moreover, the fluorescence has been successfully detected with a newly designed integrated a-Si:H detector, in which the detector and laser are placed on the same side of the microchip. This optical design is ideal for monolithic integration of the a-Si:H detector technology with VCSEL excitation.
机译:我们已经证明,氢化无定形Si(A-Si:H)光电二极管足够敏感,以用作对微生物毛细管电泳装置进行化学和生物测定的探测器。利用共焦光学观察到在50μm深微通道中流动的荧光素56μm的检测限。此外,用新设计的集成A-Si:H检测器成功地检测到荧光,其中检测器和激光放置在微芯片的同一侧。该光学设计是具有VCSEL激励的A-Si:H检测器技术的单片集成的理想选择。