Measurement of radon concentration in water using direct dpm method of liquid scintillation counter

机译:液体闪烁计数器直接DPM法测量水中氡浓度的测量

获取原文

摘要

For measuring radon (~(222)Rn) concentration in water with a direct method, a two-layer sample is measured by a liquid scintillation counter (LSC). A calibration factor for an LSC is generally evaluated by a comparison measurement of a radium (~(236)Ra) standard. However, it is difficult to use the ~(236)Ra standard in Japan. Thus, we studied a direct dpm method based on the integral counting method for a two-layer sample. The optimized window values of three efficiency tracing points for the direct dpm method were determined to be 50-2000, 75-2000, and 100-2000 keV. It was determined that a calibration factor of 5 could be used with an LSC.
机译:为了用直接方法测量水中的氡(〜(222)RN)浓度,通过液体闪烁计数器(LSC)测量双层样品。 LSC的校准因子通常通过镭(〜(236)Ra)标准的比较测量来评估。但是,难以在日本使用〜(236)RA标准。因此,我们研究了一种基于两层样本的积分计数方法的直接DPM方法。直接DPM方法的三个效率跟踪点的优化窗口值确定为50-2000,75-2000和100-2000keV。确定可以与LSC一起使用5的校准系数。